GTMG3078P350VV-X.X-FW Open Frame/Internal Medical Power Supply
GTMG3078P350VV-X.X-FW | ||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
GTMG3078P350VV-X.X-FW, Medical Power Supply, Open Frame/Internal, Open Frame Medical Power Supply, , Input Rating: 90 ~ 264VAC, 127 ~ 370VDC (47 ~ 63Hz), 3 Pin connector Mating Housing: JST VHR-3N or equivalent Mating Contact: JST SVH-21T-P1.1 or equivalent, Output Rating: 350 Watts, Power rating with convection cooling (W) , 12-48V in 0.1V increments, Approvals: WEEE; CE; China RoHS; RoHS; GOST-R; PSE; cUR; Ukraine; Double Insulation; VCCI; cULus; | ||||||||||||||||||||||||||||||||||||
Information | ||||||||||||||||||||||||||||||||||||
Model Number | GTMG3078P350VV-X.X-FW | |||||||||||||||||||||||||||||||||||
Agency Documents | http://www.globtek.info/certs/GTMG3078PWWWVV-X.X-FW/ | |||||||||||||||||||||||||||||||||||
CE EC-Declaration for Medical | Contact GlobTek for more information | |||||||||||||||||||||||||||||||||||
CE EC-Declaration for Industrial/Commercial/ITE | https://cs.globtek.info/ec_declaration?id=a0Oa000000MEcs9EAD | |||||||||||||||||||||||||||||||||||
CE EC-Declaration for Household | Contact GlobTek for more information | |||||||||||||||||||||||||||||||||||
UKCA-Declaration for Medical | Contact GlobTek for more information | |||||||||||||||||||||||||||||||||||
UKCA-Declaration for Industrial/Commercial/ITE | https://cs.globtek.info/ukca_declaration?id=a0Oa000000MEcs9EAD | |||||||||||||||||||||||||||||||||||
UKCA-Declaration for Household | Contact GlobTek for more information | |||||||||||||||||||||||||||||||||||
Morocco SDoC Declaration | https://cs.globtek.info/Morocco_SDoC_Declaration?id=a0Oa000000MEcs9EAD | |||||||||||||||||||||||||||||||||||
RoHS/RoHS3 Declaration | https://www.globtek.com/pdf/rohs_cert/a0Oa000000MEcs9EAD | |||||||||||||||||||||||||||||||||||
Prop 65 Statement | https://www.globtek.com/r2/PH7CNf | |||||||||||||||||||||||||||||||||||
REACH Declaration | https://www.globtek.com/r2/eM2co8 | |||||||||||||||||||||||||||||||||||
Conflict Minerals Declaration | https://www.globtek.com/r2/vZY8R5 | |||||||||||||||||||||||||||||||||||
Model Parameters | ||||||||||||||||||||||||||||||||||||
Type | Open Frame/Internal | |||||||||||||||||||||||||||||||||||
Technology | Open Frame Medical Power Supply | |||||||||||||||||||||||||||||||||||
Category | Medical Power Supply | |||||||||||||||||||||||||||||||||||
Input Voltage Range (V) | 90 ~ 264VAC, 127 ~ 370VDC (47 ~ 63Hz) | |||||||||||||||||||||||||||||||||||
Max Input Current (A) | 4.0A / 115VAC, 2.0A / 230VAC | |||||||||||||||||||||||||||||||||||
Max Output Power Rating (W) | 350 | |||||||||||||||||||||||||||||||||||
Output Voltage Range (V) | 12-48 | |||||||||||||||||||||||||||||||||||
Efficiency | N/A | |||||||||||||||||||||||||||||||||||
Ingress Protection | IPX0 | |||||||||||||||||||||||||||||||||||
Size (mm) | 76.2x39.6x127 | |||||||||||||||||||||||||||||||||||
Enclosure | ||||||||||||||||||||||||||||||||||||
Rating Table | ||||||||||||||||||||||||||||||||||||
* (Final ordering #/Part No to be defined) | ||||||||||||||||||||||||||||||||||||
Specifications | ||||||||||||||||||||||||||||||||||||
Features: Universal AC input / Full range with active PFC 5"x 3" footprint Medical safety approval: ANSI / AAMI ES 60601-1 / TUV EN 60601-1; IEC 60601-1 / 2xMOPP 200W convection cooling and 350W with 23.5 CFM forced air Protection: Short circuit, Over load, Over voltage High operation temperature up to 70°C Built-in fan auxiliary Power Power OK signal Remote ON / OFF, Remote sense function Withstand 2G vibration test ELECTRICAL SPECIFICATIONS: Output DC Voltage Range: 12V 15V 24V 48V Rated Current: 29.17A 23.34A 14.6A 7.3A Current Range (Convection): 0 ~ 16.67A 0 ~ 13.34A 0 ~ 8.34A 0 ~ 4.17A Current Range (23.5 CFM Fan): 0 ~ 29.17A 0 ~ 23.34A 0 ~ 14.6A 0 ~ 7.3A Rated Power (Convection): 200W 200.1W 200.1W 200.1W Rated Power (23.5 CFM Fan): 350W 350.1W 350.4W 350.4W Ripple & Noise (Max.): 150mVp-p 150mVp-p 240mVp-p 480mVp-p Voltage Adj. Range: ±5% Voltage Tolerance: ±2.0% ±2.0% ±2.0% ±2.0% Line Regulation: ±1.0% ±1.0% ±1.0% ±1.0% Load Regulation: ±1.0% ±1.0% ±1.0% ±1.0% Setup, Rise Timev 800ms, 150ms / 230VAC at full load Hold Up Time: >5ms at 350W / 230VAC Input Voltage Range: 90 ~ 264VAC, 127 ~ 373VDC Frequency Range: 47 ~ 63Hz Efficiency (Typ.) at 230VACv 90% 90% 90% 91% AC Current (Typ.): 4.07A / 100VAC, 1.64A / 240VAC Inrush Current (Typ.): Cold Start 40A / 115VAC, 80A / 230VAC Leakage Current: For earth < 300uA / 264VAC, For patient < 200uA / 264VAC Protection Over Current / Short Circuitv Constant current limit: > 105% rated power Auto-recovery: If O/P drop to 40% of the rated output voltage, PSU will shut down and auto-recover 5 times (If fault condition remains after 5 times recovery, PSU will shut down. User must re-power on to recover) Over Voltage: 115 ~ 140% rated output voltage Protection type: latch-off mode Environment Working Temp.: -20 ~ +70°C (Refer to output load de-rating curve) Working Humidity: 20 ~ 90% RH non-condensing Storage Temp. & Humidity: -40 ~ +85°C, 10 ~ 95% RH Temp. Coefficient: ±0.03% / °C (0 ~ 50°C) Vibration: 10 ~ 500Hz, 2G 10min. / 1cycle, period for 60min. each along X, Y, Z axes Safety & EMC Safety Standards Meets UL / cUL (ANSI / AAMI ES 60601-1; CAN / CSA-C22.2 No.60601-1); NFPA 99, Meets TUV (EN 60601-1; IEC 60601-1); MOPP approval Withstand Voltage I/P-O/P: 4KVAC, I/P-FG: 1.8KVAC, O/P-FG: 1.5KVAC0.5KVAC Isolation Resistance I/P-O/P: 100M Ohms / 500VDC EMI Conduction & Radiation Certified EN 55011 Harmonic Current Certified EN 61000-3-2; EN 61000-3-3 EMS Immunity Certified Certified EN 60601-1-2; IEC 61000-4-2, 3, 4, 5, 6, 8, 11 Others MTBF: XXX K HRS Certified MIL-HDBK-217F Cooling: Free Air convection for 200W, with 23.5 CFM Fan for 350W Dimension (WxHxH): 76.2x39.6x127 mm / 3.00x1.56x5.00 inch Packing: 0.42 kg Note 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25°C of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uF & 47uF parallel capacitor. 3. Tolerance: includes set up tolerance, line regulation and load regulation. 4. Line regulation is measured from low line to high line at rated load. 5. Load regulation is measured from 0% to 100% rated load. 6. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. | ||||||||||||||||||||||||||||||||||||
Derating Curve | ||||||||||||||||||||||||||||||||||||
Qty (ea.)/Carton | TBD | |||||||||||||||||||||||||||||||||||
Individual Box/Gift Box Size (mm) | 130*77*45 | |||||||||||||||||||||||||||||||||||
Individual Box/Gift Box Color | White | |||||||||||||||||||||||||||||||||||
Master Carton Size (mm) | TBD | |||||||||||||||||||||||||||||||||||
Gross Weight (kg) | TBD | |||||||||||||||||||||||||||||||||||
Label Width (mm) | TBD | |||||||||||||||||||||||||||||||||||
Label Height (mm) | TBD | |||||||||||||||||||||||||||||||||||
Input Configuration | ||||||||||||||||||||||||||||||||||||
Description | 3 Pin connector Mating Housing: JST VHR-3N or equivalent Mating Contact: JST SVH-21T-P1.1 or equivalent | |||||||||||||||||||||||||||||||||||
Approvals | ||||||||||||||||||||||||||||||||||||
Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice. Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation. | ||||||||||||||||||||||||||||||||||||
|